5.3.4.8. Preset (section 4.2.3.6)
No preset is required. Set to 0.000e+0
Relay Output 1 Setup
int.
elay Output 2 Setup
5.3.4.9. Source (section 4.2.4.1)
Navigate back to Relay Output 1 setup. Select Input A for source.
5.3.4.10. Latch Mode (section 4.2.4.2.1)
Select Off
5.3.4.11. On / Off Delay (section 4.2.4.2.2)
No delay is desired. Enter 0.00 for both.
5.3.4.12. Output Switching (section 4.2.4.4.1)
Relay one is the first of two displacement alarms to be set and will
monitor when the stage has moved greater than 4 inches in the positive
direction.
• Select EA (energize above)
• Enter 3.750e+0 for the Safe setpoint.
• Enter 4.000e+0 for the Alarm setpo
R
• Enter –3.750e+0 for the Safe setpoint.
Navigate to Relay Output 2 setup. Make all constants the same except for Output
switching.
5.3.4.13. Output Switching (section 4.2.4.4.2)
• Select EB (energize below)
• Enter –4.000e+0 for the Alarm setpoint
Digital Output 1 & 2 and Analog Output Setup
None of these functions are being used and Source for each should be set to None.
curity SetupSe
5.3.4.15. Keypad Lock (section 4.2.6.3)
5.3.4.14. Alarm Hold-Off (section 4.2.6.2)
Set to Off.