From the Analog Output SETUP pages previously
described, use
to navigate to Min Value. To
change the constant depress
while it is highlighte
The sign preceding the base number is highlighted
d.
. Use
or to change the sign. Depress to navigate the cursor to the first digit to be
changed for this example. Use or to increm
required. Use
ent or decrement the number as
for each additional digit to change. Change the exponent sign in the
e manner as described for the sign preceding the base number. Depress sam
to exit
se
changes to the constant.
U
to navigate to Max Value. Use the same
equence to change this constant.
When complete, save
han d in section 4.1.2.
Resolution is improved by keeping the span between Min and Max
Values to a minimum. Always pick values to closely match the
application.
4.2.5.2. Analog Output Calibration (TACHLINK ONLY)
Each tachometer instrument comes calibrated from the factory. In the
event the user wishes to check calibration or tailor it to a specific loop
resistance, calibration can be accomplished through TACHLINK.